• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 3256
Medical Device Problem Codes Mechanical Problem (1384); Output Problem (3005); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that rotation speed was unstable.A 1.50mm rotapro and a rotapro console were selected for use in a percutaneous coronary intervention to treat coronary artery disease.During the procedure, the speed was set to 160,000 rotations per minute (rpm), but the burr was not maintaining a consistent speed.The speed would fall below 160,000 rpm and jump over 200,000 rpm.While in dynaglide mode during the first ablation, an intermittent jumping noise was heard from the advancer, and the device would stall.Connections were checked, but the 1.50mm rotapro was ultimately exchanged for another.The errors resolved, and the procedure was successfully completed.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11449405
Report Number2134265-2021-03071
Device Sequence Number4049020
Product Code MCX
UDI-Device Identifier08714729975076
UDI-Public08714729975076
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3256
Device Catalogue Number3256
Device Lot NumberRP000836
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/15/2021
Initial Report FDA Received Date03/09/2021
Patient Sequence Number1
-
-