Model Number 71420134
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/12/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a chronic infection it was decided to explanted the implants, gii oxinium fem size 4 left, gns ii cmt tib size 5 left and lgn xlpe dished isrt sz 5-6 9mm.Site was washed out.Once wound was opened samples were taken and sent to histology.A cocr genii femur and all poly tibia was implanted.Patient outcome is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that per complaint details, the revision was performed due to an unspecified chronic infection.It was communicated that the requested medical documentation was not available for evaluation and the explanted components were not returning.¿the surgeon does not blame the implants for the issue¿.The root cause of the reported chronic infection remains unknown.The patient impact beyond the reported infection and subsequent revision could not be determined.No further medical assessment could be rendered at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.A review of sterilization documents indicated that the product was sterilized according to sterilization release documentation.Infection, a potential complication associated with any surgery, can occur and possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OXINIUM FEM SIZE 4 LEFT
Common Device Name
PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11449977
Report Number1020279-2021-01949
Device Sequence Number3513548
Product Code LZO
UDI-Device Identifier03596010430144
UDI-Public03596010430144
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/05/2021
Device Model Number71420134
Device Catalogue Number71420134
Device Lot Number11CM15562
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/12/2021
Supplement Date Received by Manufacturer07/01/2021
Initial Report FDA Received Date03/10/2021
Supplement Report FDA Received Date07/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-