US SURGICAL PUERTO RICO TA; STAPLE, IMPLANTABLE Back to Search Results
Model Number TA6035L
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, staples did not come out.There was no patient involvement.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the loading unit had been partially applied.Functionally, the instrument and the single use loading unit (sulu) were applied to appropriate test media.The jaw closed smoothly, the pin slide advanced fully, and the handle actuated smoothly into pre-clamped and clamped positions.The jaw opened, handle unlocked, and pin slide retracted when black release button was depressed.Acceptable staple formation was observed on test media.It was reported that the staples did not deploy from the stapler after the handle was squeezed.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when the firing stroke is not completed during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the instrument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The black release button may be used to open the instrument at any time during approximation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, staples did not come out.The device was exchanged to resolve the issue.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key11453062
Report Number2647580-2021-00868
Device Sequence Number14397007
Product Code GDW
UDI-Device Identifier10884521043206
UDI-Public10884521043206
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Model NumberTA6035L
Device Catalogue NumberTA6035L
Device Lot NumberP6F0017X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2021
Initial Date Received by Manufacturer 02/16/2021
Supplement Date Received by Manufacturer03/26/2021
Initial Report FDA Received Date03/10/2021
Supplement Report FDA Received Date04/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-