| Model Number |
N/A |
| Medical Device Problem Codes |
Fracture (1260); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medical devices: femur trabecular metal cruciate retaining (cr) standard porous catalog#: 42502806401 lot#: 64635133; articular surface fixed bearing cruciate retaining (cr) left 10 mm height catalog#: 42502806401 lot#: 64635133.The device will not be returned for analysis, as its current location is unknown; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent a left knee revision approximately four months post-implantation due to tibial loosening.No bone in growth was seen on the explanted tibia component.No further information is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of photograph.Visual examination of the provided picture identified evidence of being implanted and a fractured peg.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent a left knee revision approximately four months post-implantation due to tibial loosening.No bone in growth was seen on the explanted tibia component.The bone had noted osteosclerosis.The tibial tray peg fractured during removal.
|
| |
|