| Model Number |
00884450160503 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Air Embolism (1697)
|
| Date of Event |
02/24/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a percutaneous coronary intervention [pci], air was accidentally injected into the patient's vasculature system.The physician had successfully acquired arterial access and had negotiated the patient's coronary artery under fluoroscopy with a guidewire and an interventional guide catheter.During the procedure, the physician was complaining about a hemostasis valve that was leaking fluid and continued to use the hemostasis device.During the procedure air was accidentally injected into the patient's coronary artery, ecg changes were noted and conformed that the patient was experiencing a myocardial infarction [mi] and the procedure was stopped to observe vital signs and hemodynamics.Acls and basic life supporting protocols were activated and successfully administered per hospital policy.Additional lifesaving interventions, techniques, and medications were required.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The device was examined visually and microscopically.Functional testing was performed.Upon completion of the inspection no issues were observed.No leaks observed during investigation and the unit met inspection criteria.The complaint could not be confirmed.No definitive root cause could be identified upon completion of the investigation.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
| |
|
Search Alerts/Recalls
|