| Model Number |
NN801Z |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Additional information investigation results will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap ag that a columbus knee system (part # n801z) was implanted during a procedure performed on (b)(6), 2021.According to the complainant, before unpacking the columbus as implant, a coating issue was observed.Reportedly, the cement pockets appeared to be without as coating.The procedure was completed using a replacement device.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the event.Although requested, additional information has not been made available.The malfunction is filed under aag reference cc (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: visual investigation: the components were examined visually and microscopically.The surface of the implant (area to which the bone cement is applied) shows visible discoloration.The discoloration is not a deviation of the material but oxidation.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a no failure of the coating of the implant was detected.There is no indication for a material-, manufacturing- or design-related failure.The discoloration is not a deviation of the material but oxidation.Such changes in color have no effect on the function or properties of the coating.This is also mentioned in the ifu.Based upon the investigations results a capa is not necessary.
|
| |
|
Search Alerts/Recalls
|