• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US GLOBAL AP EXTRACTION HANDLE; EXTREMITY INSTRUMENTS : HANDLES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US GLOBAL AP EXTRACTION HANDLE; EXTREMITY INSTRUMENTS : HANDLES Back to Search Results
Model Number 2130-01-075
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The base of the handle of the stem inserter became loose, allowing it to slide down the shaft of the inserter, after the surgeon hammered it (in the correct location) in an attempt to remove a stem.This occurred during a revision case where a global ap was being removed.The surgeon was able to use other instruments within our shoulder extraction trays to successfully remove the stem and complete the operation successfully.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : examination of the returned device confirmed the reported event.The noted damage is consistent with device wear out and the investigation did not establish a need for corrective action.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL AP EXTRACTION HANDLE
Common Device Name
EXTREMITY INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11457144
Report Number1818910-2021-04888
Device Sequence Number3597514
Product Code LXH
UDI-Device Identifier10603295090588
UDI-Public10603295090588
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2130-01-075
Device Catalogue Number213001075
Device Lot NumberAF0308
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2021
Initial Date Received by Manufacturer 03/02/2021
Supplement Date Received by Manufacturer06/01/2021
07/30/2021
Initial Report FDA Received Date03/10/2021
Supplement Report FDA Received Date06/11/2021
08/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-