• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PTS DIAGNOSTICS CARDIOCHEK PLUS ANALYZER; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PTS DIAGNOSTICS CARDIOCHEK PLUS ANALYZER; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number 2702
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the analyzer started overheating upon replacing the batteries.There was no allegation of patient/user harm.There was no allegation of incorrect results.Currently it is unknown whether or not the device may have malfunctioned, as the allegation could not be duplicated.It is known that improper insertion of batteries - in any device - may result in overheating, battery leakage, etc., which is stated in the labeling of most battery manufacturers.
 
Event or Problem Description
The customer reported that the analyzer started overheating upon replacing the batteries.There was no allegation of patient/user harm.There was no allegation of incorrect results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOCHEK PLUS ANALYZER
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
PTS DIAGNOSTICS
4600 anson blvd
whitestown IN
MDR Report Key11461895
Report Number1836135-2021-00005
Device Sequence Number3997494
Product Code CGA
UDI-Device Identifier00381932702010
UDI-Public(01)00381932702010(8012)111
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
140068
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2702
Device Catalogue Number2702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2021
Initial Report FDA Received Date03/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-