| Model Number |
1012449-08 |
| Medical Device Problem Codes |
Material Separation (1562); Failure to Advance (2524); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional device referenced is filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a de novo lesion in the moderately calcified, moderately tortuous left anterior descending coronary artery.After an unspecified stent was implanted, a 3x8mm nc trek balloon dilatation catheter (bdc) failed to cross the lesion due to the anatomy, and the proximal shaft separated into two pieces.The bdc was simply withdrawn without resistance.A 2.75x8mm nc trek bdc also failed to cross the lesion due to the anatomy, and the proximal shaft separated into two pieces.The bdc was simply withdrawn without resistance.Then a 2.5x8mm nc trek bdc was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional and functional inspections were performed on the returned device.The reported separation and tear were confirmed.The reported failure to advance could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a de novo lesion in the moderately calcified, moderately tortuous left anterior descending coronary artery.After an unspecified stent was implanted, a 3x8mm nc trek balloon dilatation catheter (bdc) failed to cross the lesion due to the anatomy, and the proximal shaft separated into two pieces.The bdc was simply withdrawn without resistance.A 2.75x8mm nc trek bdc also failed to cross the lesion due to the anatomy, and the proximal shaft separated into two pieces.The bdc was simply withdrawn without resistance.Then a 2.5x8mm nc trek bdc was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Subsequent to the initially filed report, the following information was received: during prep, the 2.75x8mm nc trek balloon was slightly inflated to check for any leak.A tear in the guide wire exit notch of the 3x8mm nc trek balloon was noted.No additional information was provided.
|
| |
|