• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ELITE PLUS LPW CUP 28/53; MISCELLANEOUS HIP CUP/LINER : HIP POLY ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ELITE PLUS LPW CUP 28/53; MISCELLANEOUS HIP CUP/LINER : HIP POLY ACETABULAR CUPS Back to Search Results
Model Number 9651-28-053
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 02/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Original total hip replacement (c-stem) was conducted (b)(6) 2005 at (b)(6) hospital.Today ((b)(6) 2021), the elite plus lpw cup was removed and a competitor cup implanted.The original elite head was also removed and a new one implanted.Patient demographics: unknown.Reason for revision: instability.Further patient details and operation notes unavailable.
 
Event or Problem Description
Additional information received indicated that the affected side was the left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : 45 parts were manufactured per specification.3 parts were scrapped due to marred faces.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : a manufacturing record evaluation was performed for the finished device yem-01 number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot a manufacturing record evaluation was performed for the finished device yem-01 number, and no non-conformances / manufacturing irregularities related to the malfunction were identified.45 parts were manufactured per specification.3 parts were scrapped due to marred faces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELITE PLUS LPW CUP 28/53
Common Device Name
MISCELLANEOUS HIP CUP/LINER : HIP POLY ACETABULAR CUPS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key11463759
Report Number1818910-2021-04950
Device Sequence Number4188790
Product Code JDI
UDI-Device Identifier10603295318569
UDI-Public10603295318569
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9651-28-053
Device Catalogue Number965128053
Device Lot NumberYEM-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/25/2021
Supplement Date Received by Manufacturer05/05/2021
12/14/2021
05/17/2022
Initial Report FDA Received Date03/11/2021
Supplement Report FDA Received Date05/06/2021
12/15/2021
05/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ELITE MODULAR HD 28MM +3; ELITE PLUS LPW CUP 28/53
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-