| Model Number |
13115535 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Pneumothorax (2012)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No radiographs were provided for review.No device was returned as the device was not believed at fault in the event.Review of surgeons notes identified he made a mis-selection of the screw length and had an error in creating the pilot hole.The patient has recovered from pneumothorax and is doing well.".Potential risks identified with the use of this system, which may require additional surgery, include:.Neurological, vascular or visceral injury ¿" ".Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.".
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021 a male patient underwent a spinal procedure.During the procedure it was reported that a pedicle screw penetrated the anterior wall of the vertebral body at right th3 and contacted the patients lungs, resulting in pneumothorax.The screw was removed and the physician installed a chest tube and resumed the operation.The surgeon believed he made a mis-selection of the screw length and had an error in creating the pilot hole.The patient has recovered from pneumothorax and is doing well.
|
| |
|