• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TANDEM STR THREADED HNDLE; PROSTHESIS,HIP,FEMORALCOMPONENT,CEMENTED,METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TANDEM STR THREADED HNDLE; PROSTHESIS,HIP,FEMORALCOMPONENT,CEMENTED,METAL Back to Search Results
Model Number 71362804
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a tandem bipolar surgery, the tandem str threaded hndle broke, instruments outside patient.Procedure finished with s&n backup device.No surgical delay or injury to patient reported due this event.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, the product analysis and the reported event could not be confirmed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TANDEM STR THREADED HNDLE
Common Device Name
PROSTHESIS,HIP,FEMORALCOMPONENT,CEMENTED,METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11464463
Report Number1020279-2021-01994
Device Sequence Number3618932
Product Code JDG
UDI-Device Identifier03596010497994
UDI-Public03596010497994
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362804
Device Catalogue Number71362804
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 02/14/2021
Supplement Date Received by Manufacturer05/28/2021
Initial Report FDA Received Date03/11/2021
Supplement Report FDA Received Date05/31/2021
Patient Sequence Number1
-
-