MAUDE Adverse Event Report: ZIMMER BIOMET, INC. RGX 3 PEG SER A PATELLA 31MM; PROSHTESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. RGX 3 PEG SER A PATELLA 31MM; PROSHTESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Loss of or Failure to Bond (1068); Fracture (1260); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Scar Tissue (2060); Joint Laxity (4526)
Date of Event 11/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Medical product vngd ps open por fmrl rt 67.5 item# 184510 lot# 494730; bmet regenx pri tib tray 71mm item# 141273 lot# 360190; biomet finned pri stem 40mm item# 141314 lot# 943150; rgx 3 peg ser a patella 31mm item# 141356 lot# 870260; e1 vngd ps+ tib brg 71/75x14 item# ep-183744 lot# 575950.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent a right knee revision procedure approximately five years¿ post-implantation due to pain, implant fracture, scar tissue, and loosening.There is no additional information available at this time.
 
Additional Manufacturer Narrative
(b)(4).Reported event was confirmed per the medical records which found a broken patella component causing pain.A fibroma was excised.A lot of scar tissue was found beneath the patella.The patella was found loose and when removed all 3-pegs were found broken off and remained in the patella.Device history record was reviewed and no discrepancies relevant to the reported event were found.The root cause of the event was determined to be a labeling and training deficiency as the surgical technique addendum for installation of regenerex patella failed to include guidance on depth line indication on the peg drill bit.This event was farther investigated through the capa process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
It was further reported that the patient had a right total knee arthroplasty.Subsequently, two years post-op, the patient fell and began to have stiffness and pain in the patella region.The patient reported that the knee was giving out and the patient continued with pain until they were subsequently revised 5 years post-op.During the revision, a fibroma was excised, extensive scar tissue beneath the patella was encountered, the patella was loose, and three pegs of the patella were found fractured.All available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; b6; d2; g1; g3; g6; h1; h2; h6.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h6.Additional medical records were provided.Notes reported stiffness and pain coming from patella after a fall.X-ray was taken and did not show abnormalities.Office visit notes states that patient's knee continued giving pain and giving out and was treated with steroid injections.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RGX 3 PEG SER A PATELLA 31MM
Common Device Name
PROSHTESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11464660
Report Number0001825034-2021-00732
Device Sequence Number3191453
Product Code MBH
UDI-Device Identifier00880304554412
UDI-Public(01)00880304554412(17)200901(10)870260
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K083782
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2020
Device Model NumberN/A
Device Catalogue Number141356
Device Lot Number870260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/18/2021
Supplement Date Received by Manufacturer04/09/2021
08/15/2023
09/15/2023
Initial Report FDA Received Date03/11/2021
Supplement Report FDA Received Date04/09/2021
09/11/2023
09/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age49 YR
Patient SexFemale
Patient Weight159 KG
-
-