| Model Number |
5271308P2 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hematoma (1884); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The involved nuvasive implants remain in-situ as no product malfunction was suggested or identified.No radiographs were provided, photographs confirmed metal fragment.The severity and root cause of the hematoma could not be confirmed but is suggestively the result of inadvertent contact from an implant or instrument during the index procedure.The metal fragment found and removed during the revision procedure could not be identified and the origin is unknown.The piece has been retrieved and the patient has recovered well post revision.No additional investigation can be completed.".Potential adverse events and complications.As with any major surgical procedures, there are risks involved in spinal/orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include:.Damage to blood vessels.Tissue reactions including macrophage and foreign body reactions adjacent to implants.".
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021 a patient underwent an anterior spinal procedure to address stenosis at the l5-s.On (b)(6) 2021 a revision surgery was performed to remove a discovered hematoma at the left pedicle at l5/s.During the revision a metal piece was discovered inside the patient when the surgeon made an incision to remove the hematoma.There was no damage or adverse effects to patient as a result of the recovered fragment and it was retrieved during the revision.
|
| |
|
Search Alerts/Recalls
|