• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Physical Entrapment (2327)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted to report study findings reported in the literature article as well as investigation results.The device(s) referenced in this report have not been returned to olympus for evaluation, therefore physical evaluation of the complaint device could not be performed.The device history record (dhr) for the complaint device could not be reviewed since the serial number was not provided.Olympus does not ship any device that does not meet all design and safety specifications conclusion: no device malfunction was reported by the authoring physician.The physician attributes the adverse event a rare procedural complication that can occur as a result of treating of large proximal ureteral stones.
 
Event or Problem Description
It is reported in the literature article titled "between a rock and a hard place: a case report of stone fragment impaction causing a retained ureteroscope requiring open surgical intervention", a patient experienced physical entrapment of the complaint device requiring surgical intervention.As reported in the case study, a (b)(6)-year-old female underwent ureteroscopy for a 2 cm right ureteropelvic junction obstructing stone.After laser lithotripsy, there was significant buildup of stone debris distally along the ureteroscope.Conservative measures failed to remove the ureteroscope.The entrapment of the device is reported to be due to heavy stone burden surrounding the tip of the device preventing it's safe withdrawal.An exploratory laparotomy through a two sides gibson incision and right ureteral dissection for manual removal of the stone debris was performed with eventual removal of the retained ureteroscope.A ureteral reimplant was performed.Postoperative computerized tomography (ct) showed residual hydronephrosis, but there was no obstruction seen on renal lasix scan.There is no report of device malfunction.The study concluded that ureteroscopic treatment of large proximal ureteral stones may lead to the rare complication of ureteroscope retainment caused by stone impaction.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct g2.As a result of confirmation, there are no similar complaints.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Common Device Name
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11465524
Report Number8010047-2021-03600
Device Sequence Number19740717
Product Code FGB
UDI-Device Identifier04953170287183
UDI-Public04953170287183
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K072957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberURF-V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2021
Supplement Date Received by Manufacturer01/14/2022
Initial Report FDA Received Date03/11/2021
Supplement Report FDA Received Date02/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-