| Model Number |
401753 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment.Upon removal of the stylet a deformation was detected.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes the returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the stent is severely deformed at its proximal end, i.E.Several struts are bent.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the deformed struts were initially crimped on the balloon.The balloon is well folded and shows no signs of inflation.A reference stent protector could not be slid over the deformed struts without bending them further.It can therefore be excluded beyond reasonable doubt that the damage was already present on delivery of the instrument.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|
Search Alerts/Recalls
|