It was reported that sometime during the week of (b)(6) 2021, the 3.5 x 20 mm trek balloon dilatation catheter (bdc) was advanced into the anatomy and air prepped.Contrast mix was 40/60.The device was then advanced to the heavily calcified lesion for pre-dilatation.The balloon was inflated without issue; however, after inflation, the balloon failed to deflate.Negative was held for up to a couple of minutes, but the balloon did not deflate.Resistance was felt during removal since the balloon remained inflated.There were no adverse patient effects and no reported clinically significant delay in the procedure.No additional information was provided.
|
|
The device was returned for analysis.Visual, dimensional and functional inspections were performed on the returned device.The reported deflation issue was confirmed.The reported difficulty removing the device could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.In this case, the contrast ratio used was less than the required percentage.It should be noted that the coronary dilatation catheters (cdc), trek rx and mini trek rx, global instructions for use (ifu), specifies that the contrast is 60% contrast medium diluted 1:1 with normal saline.Additionally, it was reported that the device was prepped inside the anatomy.The ifu specifies: all air must be removed from the balloon and displaced with contrast prior to inserting into the body, otherwise, complications may occur.In this case, it is unknown if the ifu violations caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported deflation issue; however, the reported difficulty removing the device appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.Na.
|