Model Number W1TR01
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 4543 lead, implant date unknown; product id: 0185 lead, implant date unknown; product id: 4136 lead, implant date unknown.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the cardiac resynchronization therapy pacemaker (crt-p) changeout procedure, after successfully connecting the existing competitor lead to the new device without issue, the physician determined the lead was not fully seated in the header.The lead was unscrewed and repositioned within the header.Upon multiple attempts to engage and tighten the set screw with multiple wrenches the physician was unable to re-secure the lead within the header.The physician then removed the grommet and attempted to tighten the setscrew, and was still unable.Finally the physician removed the setscrew and successfully used a new one without consequence.The lead was tested and a good connection with the device was confirmed.Medical adhesive was used at the site of the set screw to fill in the place where the grommet had been.The device is still in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPTA CRT-P MRI SURESCAN
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11467410
MDR Text Key239430073
Report Number9614453-2021-00994
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00643169891401
UDI-Public00643169891401
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2022
Device Model NumberW1TR01
Device Catalogue NumberW1TR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/23/2021
Initial Date FDA Received03/11/2021
Date Device Manufactured11/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-