MEDTRONIC EUROPE SARL PERCEPTA CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
|
Back to Search Results |
|
Model Number W1TR01 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 4543 lead, implant date unknown; product id: 0185 lead, implant date unknown; product id: 4136 lead, implant date unknown.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during the cardiac resynchronization therapy pacemaker (crt-p) changeout procedure, after successfully connecting the existing competitor lead to the new device without issue, the physician determined the lead was not fully seated in the header.The lead was unscrewed and repositioned within the header.Upon multiple attempts to engage and tighten the set screw with multiple wrenches the physician was unable to re-secure the lead within the header.The physician then removed the grommet and attempted to tighten the setscrew, and was still unable.Finally the physician removed the setscrew and successfully used a new one without consequence.The lead was tested and a good connection with the device was confirmed.Medical adhesive was used at the site of the set screw to fill in the place where the grommet had been.The device is still in use.No patient complications have been reported as a result of this event.
|
|