|
Model Number 71674079 |
Device Problems
Crack (1135); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, during set up or inspection two ruler failed.The rulers were excessively worn and cracked.Incident occurred before the procedure; therefore, there was no patient involvement.
|
|
Manufacturer Narrative
|
G4, h2, h6: the devices, intended for use in treatment was returned for evaluation: a visual inspection was performed and confirmed the two rulers appear to be dull and excessively worn.The manufactured date for these devices is 2019.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.D9 and h3: device returned.
|
|
Manufacturer Narrative
|
H3, h6: the associated device, intended for use in treatment, was returned, and evaluated.A visual inspection of the returned ruler confirmed the device is cracked on the tip with excessive wear usage.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|