It was reported whilst using the mobile programmer, the initial interrogation of the device was successful.The device was connected to the leads, and it was not possible to perform impedance measurements.The display showed connection between the patient connector and device but a diagnostic message was displayed on screen stating that there was no telemetry with the implantable device.The patient connector was moved around but nothing changed.The patient connector was placed above the device whilst in the patient using a sterile sleeve but the issue persisted.The application then froze and would not respond the application was closed and opened; the patient connector was able to interrogate the device.The programmer remains in use.No patient complications have been reported as a result of this event.
|
Product analysis: analysis was unable to confirm the customer comment in relation the crash.It was noted that the event was due to a known documented issue related to bluetooth.There was no report generated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|