Model Number UCR |
Device Problem
Backflow (1064)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not been returned to omsc for evaluation but was returned to sorc.Sorc checked the subject device and found that the reported phenomenon.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus service operation repair center (sorc), it was found that there was a trace of liquid ingress into the inside tube of the subject device.There was no report of patient injury associated with this event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Since the subject device was not returned to olympus medical systems corp.(omsc), it could not be investigated.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the information from olympus service operation repair center (sorc), there was the possibility that this phenomenon was attributed to the backflow of liquid because there was a trace of liquid ingress into the inside tube of the subject device.If additional information becomes available, this report will be supplemented.
|
|
Manufacturer Narrative
|
This supplemental report is to inform that upon further review, per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.Initially, we determined that the event was mdr reportable due to a backflow of liquid from the water container into the unit co2 regulator (ucr), in which case we determined that it was a potential adverse event because of the risk for infection.Upon further investigation, it was found that liquid does not flow back into the ucr from the water container unless multiple situations occur simultaneously.There are no reports of situations occurring for this complaint.In addition, a component analysis was performed on the water droplet traces in the tube of the ucr at a similar complaint.The results of this component analysis detected silica and others that appeared to be derived from tap water, but no component that appeared to be body fluids were detected.
|
|