• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LLC XPS SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LLC XPS SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884380EM
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A manufacturer representative reported that a blade vibrated at an unsteady rate during an endoscopic sinus surgery (ess).It was reseated and the blade still no change in the excessive vibration.Grabbed another blade and it functioned properly.There was no patient impact.After surgery was completed successfully, grabbed a couple more blades off their inventory shelf to perform testing and see if there was the same problem.Also observe the same excessive vibration on two other blades.
 
Additional Manufacturer Narrative
H3: analysis found the middle assembly would rotate freely.The inner assembly cutter rotated freely.The blade seated into the handpiece with ease.The blade oscillated in the lab and exhibited a wobble and vibration.The inner blade was removed and spun on a flat raised surface.The hub did not appear to be concentric.There was likely bending of the shaft resulting in a concentricity issue of the hub or concentricity issue of the hub in the molding process.H6: fdc d16 fdm b21 fdr c21 no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS SYSTEM NAVIGATED BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire street
mansfield 02048
MDR Report Key11474214
Report Number9612501-2021-00396
Device Sequence Number4537598
Product Code EQJ
UDI-Device Identifier00763000034023
UDI-Public00763000034023
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K130608
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/27/2022
Device Model Number1884380EM
Device Catalogue Number1884380EM
Device Lot Number0221182518
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2021
Supplement Date Received by Manufacturer05/12/2021
Initial Report FDA Received Date03/12/2021
Supplement Report FDA Received Date05/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age47 YR
-
-