ARTHREX, INC. ECLIPSE HUMERAL HEAD EXTRACTOR; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
| Model Number |
ECLIPSE HUMERAL HEAD EXTRACTOR |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is now reportable.During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed, the tip of the device was found to be damaged.A fragment broke off and was not returned.The cause of the event is undetermined, however the likeliest cause of the observed failure mode is user-applied mechanical forces.
|
| |
|
Event or Problem Description
|
|
It was reported that the ar-9401-17 humeral head extractor is damaged.Additional information received on 1/27/2021: the rep reported damage to the device did not happen during a case.The issue was noticed by a sterile processing tech during inspection prior to a case.This is now reportable.During returned device evaluation, a reportable malfunction was discovered.Complaint confirmed, the tip of the device was found to be damaged.A fragment broke off and was not returned.The cause of the event is undetermined, however the likeliest cause of the observed failure mode is user-applied mechanical forces.
|
| |
|