| Model Number |
C4120 |
| Medical Device Problem Codes |
Material Frayed (1262); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ra has just received the incident device and has been assigned to engineering for evaluation.A follow-up report will be sent upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: lower anterior colon resection.The account mgr was present for the case.[name] was performing a lower anterior colon resection.The grasper was being used with the ctf03 trocar.He noticed that the jaws of the grasper were misaligned.He was not grabbing tissue at the time and had already placed the drain in the patient.He stated that he did not recollect when or how the jaw became misaligned.A new grasper was opened and the case was completed.There was no damage to the tissue.There was no patient injury.There are no photos available.The device is available to be returned.Replaced with a new grasper.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed that the grasper jaws were misaligned.One of the pads had partially detached from the grasper jaw.The adhesive remnants and pad fragments on the jaw were indicative of proper adhesion and good pad attachment.Based on the condition of the returned unit, the pad detachment was caused by adhesive failure.However, there was no evidence of a manufacturing defect.As a result, the exact root cause of the pad detachment is unknown.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|
Event or Problem Description
|
|
Procedure performed: lower anterior colon resection.The account mgr was present for the case.Dr [name] was performing a lower anterior colon resection.The grasper was being used with the ctf03 trocar.He noticed that the jaws of the grasper were misaligned.He was not grabbing tissue at the time and had already placed the drain in the patient.He stated that he did not recollect when or how the jaw became misaligned.A new grasper was opened and the case was completed.There was no damage to the tissue.There was no patient injury.There are no photos available.The device is available to be returned.Intervention: replaced with a new grasper.Patient status: no patient injury.
|
| |
|