| Model Number |
SOFHDPLI+22.75D AND PLITIP1645 KIT |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.Based on the assessment of our batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the lens were conducted correctly.Lenstec believes that the lens, its design and the manufacturing process are not at fault due to extensive testing conducted on this model.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email stating "fibres have been noted either embedded into or adhered to the lens optic.These strands look similar to a crease in the lens optic yet do not dissipate".
|
| |
|
Additional Manufacturer Narrative
|
|
A full audit of all batch documentation relating to the production of the device was performed.The audit concluded that all procedures in manufacturing and packaging of the device had been conducted correctly.Batch reconciliation was 100.0%.Based on the assessment of our batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the lens were conducted correctly.Lenstec believes that the lens, its design and the manufacturing process are not at fault due to extensive testing conducted on this model.Pli lenses from the complaint batch were tested for injection integrity and no crease marks or particulate were found.
|
| |
|
Event or Problem Description
|
|
Lenstec received an email stating "fibres have been noted either embedded into or adhered to the lens optic.These strands look similar to a crease in the lens optic yet do not dissipate".
|
| |
|