| Model Number |
72203967S |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the camera head was not taking pictures.It is unknown whether the event happened during surgery and if there was patient involvement.Smith and nephew back-up device was available to complete the surgery.Unknown if there was a delay.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device, intended for use in treatment, was returned to the designated complaint unit for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.Functional evaluation revealed that the right button functions intermittently.The complaint has been confirmed and the root cause has been associated with a mechanical component failure.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event include a failed button switch.A review of the device history records showed there were no indications to suggest that the lot did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|