SMITH & NEPHEW, INC. BIOINDUCTIVE IMPLANT W/ARTH DEL LRG; MESH, SURGICAL, COLLAGEN, ORTHOPAEDICS, REINFORCEMENT OF TENDON
|
Back to Search Results |
|
| Catalog Number |
4566 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Fluid Discharge (2686)
|
| Date of Event |
10/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the regeneten bioinductive implant "scaffold" was used on a patella tendon revision.The implant was sutured on the top of the repair and no staples were used.Three months after the operation, a wash out was preformed because the patient developed a recalcitrant pre-patellar bursitis.The doctor aspirated many times and a milky fluid came out.The cultures were negative.The doctor believes that the inflammatory reaction happened due to the scaffold breaking down.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: updated data on h6 (clinical code and impact code).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined.Since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the device specifications found all dimensions or notes w/ circular number annotations will be inspected at incoming inspection.A clinical review states that the impact to the patient beyond that which has already been reported is unknown.Should any additional medical information be provided, this complaint will be re-assessed.Therefore, no further clinical/medical assessment is warranted at this time.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|