MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TAN/FAN PERC DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TAN/FAN PERC DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631021
Medical Device Problem Codes Break (1069); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during set up for a im nail procedure, the bolt from broken impactor is stuck in the threads and cant be removed.Procedure was performed with a backup device, no delay or further complications reported.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.It appears that the bolt from broken impactor is stuck in the threads and can¿t be removed rendering it inoperable.The device shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAN/FAN PERC DRILL GUIDE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11478864
Report Number1020279-2021-02058
Device Sequence Number3276281
Product Code HSB
UDI-Device Identifier03596010604873
UDI-Public03596010604873
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631021
Device Catalogue Number71631021
Device Lot Number12BM01352
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Initial Date Received by Manufacturer 02/16/2021
Supplement Date Received by Manufacturer05/28/2021
Initial Report FDA Received Date03/14/2021
Supplement Report FDA Received Date05/31/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-