Model Number 71430260
Device Problems Crack (1135); Material Discolored (1170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2021
Event Type  malfunction  
Event Description
It was reported that, spacer trials seem decolorated and cracked.No case reported; therefore, there was no patient involvement.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device has discoloration and it is cracked, rendering the device inoperable.The device was manufactured in 2004 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 INS SPAC TR SZ5-6 25MM
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11478886
MDR Text Key239690053
Report Number1020279-2021-02077
Device Sequence Number1
Product Code LXH
UDI-Device Identifier03596010212030
UDI-Public03596010212030
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71430260
Device Catalogue Number71430260
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/18/2021
Initial Date FDA Received03/14/2021
Supplement Dates Manufacturer Received02/18/2021
Supplement Dates FDA Received04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-