SMITH & NEPHEW, INC. REDAPT DISTAL REAMER 190MM SZ 14; PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT
|
Back to Search Results |
|
| Model Number |
71355345 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the end of reamer shaft broke because the doctor was using a "ronguer" to try and extract it.The neck trial will no longer attach to the reported reamer.The event occurred during use, while inside the patient.The procedure was completed using a sn backup device.No surgical delay or injury to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.Based on the alleged failure and the date when the product was manufactured, a further review of the manufacturing records was not deemed necessary.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|