• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NDI CAMERA POLARIS SPECTRA; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NDI CAMERA POLARIS SPECTRA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR200027
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2021
Event Type  Injury  
Event Description
It was reported that before a navio procedure, during the pre-op registration of camera and foot pedals, the warning "camera infrared lamp is not working properly.This may result in limited field of view" was displayed.They changed to manual procedure to complete the surgery.No other complications were reported.
 
Manufacturer Narrative
The reported device (pn 200027 sn (b)(6)) was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional evaluation could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found similar reports, this issue will continue to be monitored.This failure mode is identified in the navio risk profile.The navio user's manual (500196) provides instruction for camera setup, operation and troubleshooting.A relationship, if any, between the subject device and the reported event could not be determined.The reported error message indicates that the illuminator leds have gone bad.The illuminator leds on the camera can go bad over time with use and cause floating "dead zones" in the camera view.This problem is physical in nature in the camera only and it needs to be serviced.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NDI CAMERA POLARIS SPECTRA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11479017
MDR Text Key239690173
Report Number3010266064-2021-00187
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556649541
UDI-Public885556649541
Combination Product (y/n)N
PMA/PMN Number
K152574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberPFSR200027
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/17/2021
Initial Date FDA Received03/14/2021
Supplement Dates Manufacturer Received06/15/2021
Supplement Dates FDA Received06/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-