Model Number N/A
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Product has not been returned.
 
Event or Problem Description
It was reported tat the doctor has reported that the reamer (spherical cutter) is not performing as required, and has requested a replacement.No delay reported.No harm of the patient or the user reported.
 
Additional Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.As the product has not been received, the investigation was limited to the information provided; a review of device history records and complaint history.We have not been provided with any supporting documentation which could provide additional information.A review of the device history records did not identify any discrepancies that would have contributed to the reported event.A review of the complaint database over the last 3 years has found 3 similar complaints identified with this item number including (b)(4), and 1 similar complaint identified for this lot number including (b)(4).Should the complaint product become available, the complaint investigation will be updated as appropriate.Risk assessment: risk management report documents the estimated residual risk associated with the reported event.The root cause of the issue could not be determined with the information currently available, therefore the specific failure cause within the risk tables could not be selected for comparison.The details of the reported event do not allege that there was any patient harm or delay to surgery.The reported event states reamer not functioning properly.This incident has a severity of 1 (negligible) defined within the severity table as: no / virtually no injury to the patient or user.No impact to patient management.Insignificant or no device impairment or impact to patient (implants).Product system functionality not affected.The outcome of the reported event is considered to be within the severity of the rmf.No corrective or preventive actions are deemed necessary at this time.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported tat the doctor has reported that the reamer (spherical cutter) is not performing as required, and has requested a replacement.Excess force had to be applied for the reamer to cut.No delay reported.No harm of the patient or the user reported.
 
Additional Manufacturer Narrative
(b)(4).This follow-up final report is being submitted to relay additional information.Complaint summary: a visual check of the returned product (oxford uni spherical cutter small item (b)(4), lot.Zb151001) confirms that the cutting edges are worn on both the inner and outer parts of the cutter, the cutter otherwise appears to be in good condition.A dimensional inspection is not required for this event as this reported event is for a worn cutter.The root cause of the cutter not performing as expected, is that the product has exceeded its useful life as there is evidence of wear on all the cutting edges.This suggests the instrument has been used for many procedures since it was manufactured in 2015.The batch part is conforming before final release from the manufacturing site, and no deviations are related to the reported event.The batch part is conforming before final release from manufacturing site, no deviations are related to the reported event.A visual inspection of this device shows this event does not differ from previous reported events for the oxford uni spherical cutter small as a result there is no change to the severity or occurrence for this event with the reported event is still considered to be within the severity of the rmf.Capa: no corrective or preventive action required at this time.Investigation is completed based on current available information.If any additional information becomes available, then the complaint will be reopened and investigated.
 
Event or Problem Description
It was reported that the doctor has reported that the reamer (spherical cutter) is not performing as required, and has requested a replacement.Excess force had to be applied for the reamer to cut.No delay reported.No harm of the patient or the user reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF UNI PH3 SPH CUTTER SML ALL
Common Device Name
KNEE ARTHROPLASTY
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key11482923
Report Number3002806535-2021-00080
Device Sequence Number3412965
Product Code HXC
UDI-Device Identifier05019279482252
UDI-Public05019279482252
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number32-420329
Device Lot NumberZB151001
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2021
Supplement Date Received by Manufacturer03/23/2021
11/29/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date03/25/2021
11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-