Model Number ENO DR |
Device Problems
Pacing Inadequately (1442); Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
Reportedly, the patient is pacemaker dependent with a sinus rhythm at 70 bpm.The leads were implanted with good measurements (5mv for the atrial detection).The leads were connected to the subject pacemaker; pacing was observed in both chambers.After the intervention, the pacemaker was interrogated which terminated the automatic implantation detection.Atrial sensing test revealed p-wave amplitude values at 3.06mv; however, the pacemaker was still pacing in the atrial channel.The pacing mode was reprogrammed to vdd and the pacemaker followed then the spontaneous atrial rhythm.The device was reprogrammed to ddd mode and normal pacing following p-wave could be observed.
|
|
Manufacturer Narrative
|
D3 and g1 updated.Please refer to the attached analysis report.
|
|
Event Description
|
Reportedly, the patient is pacemaker dependent with a sinus rhythm at 70 bpm.The leads were implanted with good measurements (5mv for the atrial detection).The leads were connected to the subject pacemaker; pacing was observed in both chambers.After the intervention, the pacemaker was interrogated which terminated the automatic implantation detection.Atrial sensing test revealed p-wave amplitude values at 3.06mv; however, the pacemaker was still pacing in the atrial channel.The pacing mode was reprogrammed to vdd and the pacemaker followed then the spontaneous atrial rhythm.The device was reprogrammed to ddd mode and normal pacing following p-wave could be observed.
|
|
Search Alerts/Recalls
|