The affected device was examined by service technician and information was provided for further investigation.The investigation also considered information gained from comparable cases.Based on the reported error codes the root cause of the event was determined a deviation in the microprocessor-bus-configuration on the pcb control.Such a deviation within the electronic system will cause a software-controlled restart (reset) of the whole electronic system.The pneumatic system opens which reduces airway pressure to ambient pressure and spontaneous breathing is possible for the patient.This system-reset is combined with an audible and visible alarm of high priority.If the deviation is corrected by this reset the device will continue ventilation with last settings after not later than 12 sec.Furthermore, an issue regarding the flow measurement was reported.A deviation regarding the flow measurement will be detected by the device and alarmed accordingly.The ventilation will be continued.The customer has to act according to the troubleshooting table in the instructions for use.The flow sensor is a wear and tear part and has to be replaced if faulty.An improved version of the cable harness spirologsensor is available in case a faulty cable harness spirologsensor has to be replaced.The device reacted as specified to a detected deviation within the electronic system and performed a restart in order to remedy the issue; this restart was accompanied by an alarm.Further alarms were generated in order to alert the user of the issue regarding the flow measurement.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|