| Model Number |
7134 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that removal difficulties were encountered.The 85% stenosed target lesion was located in the middle segment of right coronary artery.A 1.50mm x 20mm emerge balloon catheter was advanced for post-dilatation of a non-bsc implanted stent.The emerge balloon catheter was fully deflated prior withdrawal.Resistance was felt during withdrawal at the lesion site.The non-bsc implanted stent migrated from the target lesion.Additional stenting was required as a result of the migrated non-bsc stent.There were no patient complications reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of the emerge balloon catheter.The hypotube, shaft, balloon and tip were microscopically and visually examined.The hypotube was separated at 64.9cm from the strain relief, and there were numerous kinks on the device.Blood was present in the guidewire lumen and the balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported device removal difficulties.As there was separation to the device and multiple kinks.
|
| |
|
Event or Problem Description
|
|
It was reported that removal difficulties were encountered.The 85% stenosed target lesion was located in the middle segment of right coronary artery.A 1.50mm x 20mm emerge balloon catheter was advanced for post-dilatation of a non-bsc implanted stent.The emerge balloon catheter was fully deflated prior withdrawal.Resistance was felt during withdrawal at the lesion site.The non-bsc implanted stent migrated from the target lesion.Additional stenting was required as a result of the migrated non-bsc stent.There were no patient complications reported and the patient's status was stable.
|
| |
|
Search Alerts/Recalls
|