| Model Number |
M00565030 |
| Medical Device Problem Codes |
Break (1069); Positioning Failure (1158); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2021 that a wallflex duodenal stent was to be implanted in the duodenum to treat a 10 cm pancreatic cancer during a duodenal prosthesis placement procedure performed on (b)(6) 2021.Reportedly, the patient's anatomy was not tortuous and was not dilated prior to stent placement procedure.During the procedure, the distal handle detached from the catheter preventing the stent to deploy.The procedure was not completed because another wallflex duodenal stent was not available.There were no patient complications reported as a result of this event.
|
| |
|