Model Number 12120
Health Effect - Clinical Codes Bacterial Infection (1735); Insufficient Information (4580)
Date of Event 02/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the customer tested a set from the same box and it came back negative for any growth.The child had 5 days of antibiotics and was collected again with no growth in the product bag.One theory, per the customer, is that the child had a small blood level of ecoli that did not grow on culture, however when cells were collected in a small concentrated level (5ml per bottle) the e.Coli was detected.Investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer reported that they collected mononuclear cell (mnc) products on a patient three separate times, and all three products tested positive for e.Coli.All three products were collected from kits that were in the same box.The first two collections were done on the patient using a femoral line.The femoral line was removed and a jugular line was placed for the third collection.The femoral line tip was not tested when it was removed.Per the customer, there was nothing unusual about the procedure.This report is being filed for the third of three procedures.The collection occurred on (b)(6) 2021.The patient was put on meropenem on the second day collection.Patient identifier 3 febis not available at this time.The disposable set is not available for return because it was discarded by the customer.
 
Event or Problem Description
The customer declined to provide patient identifier.
 
Additional Manufacturer Narrative
Investigation: a disposables complaint history search for lot 2003193130 and found no other reports of bacterial contamination.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.The customer tested a set from the box that staff had been using (same lot number) and results came back negative for any growth.Femoral vein catheters are more likely to become infected than subclavian or internal jugular vein catheters, and multilumen catheters may also have a higher rate of infection than single lumen catheters, but this may not be clinically significant since multilumen catheters are typically used for short periods of time.Source: bullard km, dunn dl.Bloodstream and intravascular catheter infections.In: holzheimer rg, mannick ja, editors.Surgical treatment: evidence-based and problem-oriented.Munich: zuckschwerdt; 2001.Available from: https://www.Ncbi.Nlm.Nih.Gov/books/nbk7008/ per internal sterility documentation, the devices terumo bct manufactures to collect, separate, and store blood products are terminally sterilized to a sterility assurance level (sal) of
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Corrected information is provided in h.10.Investigation: per literature review: escherichia coli is a gram-negative bacterium and is the most numerous aerobic commensal inhabitant of the large intestine.Certain strains cause diarrhea and all can cause infection when they invade sterile sites.Infection of the bloodstream remains a life-threatening occurrence and is most commonly associated with the presence of a central vascular catheter, but may also be associated with a gram-negative infection in other areas of the body, such as the lung, genitourinary tract, or abdomen.Approximately 30% of hospital-acquired bloodstream infections in icus in the united states are due to gram-negative organisms, although this proportion is lower when hospital-wide data are examined.Given an adequate portal of entry, almost any gram-negative organism can cause bloodstream infection; however, the most common organisms include klebsiella species, escherichia coli, enterobacter species, and p.Aeruginosa.[a.Y.Peleg et al.Hospital-acquired infections due to gram-negative bacteria.N engl j med.2010 may 13; 362(19): 1804¿1813.] according to the cdc 'guidelines for the prevention of intravascular catheter-related infections (2011)': in the united states, 15 million central vascular catheter (cvc) days (i.E., the total number of days of exposure to cvcs among all patients in the selected population during the selected time period) occur in intensive care units (icus) each year.Studies have variously addressed catheter related bloodstream infections (crbsi).These infections independently increase hospital costs and length of stay, but have not generally been shown to independently increase mortality.While 80,000 crbsis occur in icus each year, a total of (b)(4) cases of bsis have been estimated to occur annually, if entire hospitals are assessed.Root cause: a definitive root cause for the contaminated car t cell products could not be determined.It's possible the patient had a low level catheter related blood stream infection (crbsi) or a non-catheter related infection in another part of the body.Corrected investigation: "femoral vein catheters are more likely to become infected than subclavian or internal jugular vein catheters, and multilumen catheters may also have a higher rate of infection than single lumen catheters, but this may not be clinically significant since multilumen catheters are typically used for short periods of time." provided in supplement 1, is no longer applicable to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRA OPTIA
Common Device Name
SPECTRA OPTIA COLLECT SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key11484130
Report Number1722028-2021-00110
Device Sequence Number17363804
Product Code GKT
UDI-Device Identifier05020583121201
UDI-Public05020583121201
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
BK150251
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2022
Device Model Number12120
Device Catalogue Number12120
Device Lot Number2003193130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2021
Supplement Date Received by Manufacturer04/05/2021
04/07/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date04/06/2021
04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age00007 YR
Patient Weight25
-
-