MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN CR HIGH FLEX XLPE SZ 5-6 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN CR HIGH FLEX XLPE SZ 5-6 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/ Back to Search Results
Model Number 71453121
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 02/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that two years after a bilateral tka, the patient complained of instability on the left knee.Patient suffered from obesity, and had difficulty in holding the leg, assessing stability and hyperextension.A revision surgery was performed to explant a lgn cr high flex xlpe sz 5-6 9mm.The liner was replaced.Additional details have not been provided at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports the patient underwent a revision surgery/insert exchange due to complaint of instability.Per email communication, the requested surgical reports and x-rays are not available.Therefore, the patient impact beyond the revision cannot be determined.Due to the limited information provided, no further clinical assessment is warranted at this time.Should additional clinically relevant documentation become available, the clinical/medical task may be re-opened.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to fit/sizing issue, lifetime of device or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN CR HIGH FLEX XLPE SZ 5-6 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11484444
Report Number1020279-2021-02175
Device Sequence Number11620748
Product Code JWH
UDI-Device Identifier00885556033982
UDI-Public00885556033982
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71453121
Device Catalogue Number71453121
Device Lot Number18HM22654
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/23/2021
Supplement Date Received by Manufacturer07/20/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date07/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-