SYNTHES GMBH 2.0MM UNIVERSAL VARIABLE ANGLE LOCKING DRILL GUIDE; GUIDE, SURGICAL, INSTRUMENT
|
Back to Search Results |
|
| Catalog Number |
03.211.002 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Occupation: reporter is a synthes employee.Initial reporter facility name: (b)(6) hospital.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent the open reduction internal fixation surgery for the distal humeral and elbow fractures with the plate and screws.A va drill sleeve was used to drill from the support of the posterolateral plate.At that time, the drill sleeve came off a little, but it was drilled.Then, a va locking screw was inserted.The screw was not torqued and was idling without locking.The surgeon commented that the drill sleeve came off while drilling and he could not drill properly, and it caused this event.When visually checked, the screw tip came out of the bone, so the surgeon removed the screw, and a shorter 34 mm screw was inserted, but it would not lock.When the surgeon checked again, the tip of the screw was out of the bone.The surgeon drilled again and inserted 36 mm screw.The same event occurred, and a 20 mm screw was finally inserted.This 20 mm screw did not lock, but the incision was closed at the surgeon¿s discretion and the 20 mm screw remained.The surgery was completed successfully with thirty (30) minute delay.This report involves one (1) 2.0mm universal variable angle locking drill guide.This is report 6 of 6 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|