• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SKY VARIABLE S-D SCW 16MM TI; APPLIANCE,FIXATION,SPINAL INTERVERTEBRAL BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SKY VARIABLE S-D SCW 16MM TI; APPLIANCE,FIXATION,SPINAL INTERVERTEBRAL BODY Back to Search Results
Model Number 186850016
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is jnj representative.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.The implant(s) was not returned, and the investigation will be completed based on the supplied image(s).The image(s) was reviewed, and the complaint condition of broken was confirmed as the image(s) showed several broken screws along the spine.As the implant(s) was not returned and as received, dimensional, material or drawing reviews are not applicable.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image.There is no indication that a design or manufacturing issue contributed to the complaint as the circumstances during the time of the event are unknown.No new malfunctions were observed during this investigation (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent for a removal surgery for a skyline plate/screws as the screws had fractured.The screws had fractured in the vertebral body and metals was subsequently left in the patient.The surgery was completed successfully.The patient outcome was unknown.Concomitant device reported: bone vertigraft cervical trapezoid 5*7mm (part# s18155241049, lot# unknown, quantity 1).Bone vertigraft cervical trapezoid 5*7mm (part# s17197051079, lot# unknown, quantity 1).Bone vertigraft cervical/trap fd 6*8mm (part# s1816613-1034, lot# unknown, quantity 1).This complaint involves nine (9) devices.This report is for (1) sky variable s-d scw 16mm ti.This report is 7 of 8 for (b)(4).
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: investigation summary: investigation flow: damage.Visual inspection: the sky variable s-d scw 16mm ti (p/n: 186850016, lot#: unk) was returned and received at us cq.Upon visual inspection, the screw is broken which is consistent with the complaint condition.The screws were not able to disassemble from the plate and the broken pieces were not returned.The device failure / defect was identified.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Furthermore, the complaint relevant dimensions cannot be checked for dimensional accuracy, because of the design of the device.Document/specification review: based on the date of manufacture, the current and manufactured revision of drawings were reviewed.Investigation conclusion: the complaint condition is confirmed for the sky variable s-d scw 16mm ti (p/n: 186850016, lot#: unk).There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: null, lot unknown.Device history batch: null, null.Device history review: a manufacturing record evaluation could not be performed as the lot number could not be determined from the image, null.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SKY VARIABLE S-D SCW 16MM TI
Common Device Name
APPLIANCE,FIXATION,SPINAL INTERVERTEBRAL BODY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11486088
Report Number1526439-2021-00444
Device Sequence Number14358771
Product Code KWQ
UDI-Device Identifier10705034134109
UDI-Public10705034134109
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K052552
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186850016
Device Catalogue Number186850016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2021
Initial Date Received by Manufacturer 02/16/2021
Supplement Date Received by Manufacturer04/14/2021
08/23/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date05/10/2021
08/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BONE VERTIGRAFT CERVICAL TRAPEZOID 5*7MM.; BONE VERTIGRAFT CERVICAL TRAPEZOID 5*7MM.; BONE VERTIGRAFT CERVICAL/TRAP FD 6*8MM.; SKY 3 LEVEL PLTE 54MM TI.; SKY VARIABLE S-D SCW 16MM TI.; SKY VARIABLE S-D SCW 16MM TI.; SKY VARIABLE S-D SCW 16MM TI.; SKY VARIABLE S-D SCW 16MM TI.; SKY VARIABLE S-D SCW 16MM TI.; SKY VARIABLE S-D SCW 16MM TI.; SKY VARIABLE S-D SCW 16MM TI.; BONE VERTIGRAFT CERVICAL TRAPEZOID 5*7MM; BONE VERTIGRAFT CERVICAL TRAPEZOID 5*7MM; BONE VERTIGRAFT CERVICAL/TRAP FD 6*8MM; SKY 3 LEVEL PLTE 54MM TI; SKY VARIABLE S-D SCW 16MM TI; SKY VARIABLE S-D SCW 16MM TI; SKY VARIABLE S-D SCW 16MM TI; SKY VARIABLE S-D SCW 16MM TI; SKY VARIABLE S-D SCW 16MM TI; SKY VARIABLE S-D SCW 16MM TI; SKY VARIABLE S-D SCW 16MM TI
Outcome Attributed to Adverse Event Required Intervention;
-
-