BAXTER HEALTHCARE CORPORATION INTERLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 1W5000
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an interlink system - solution set incorrectly assembled.During an albumin infusion, a new tubing set was needed.After the tubing was primed, the nurse attempted to insert it into an unspecified infusion pump; however, the set could not be loaded into the pump due to the ¿blue key on the tubing was backwards¿.The albumin was infused by gravity.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection was performed using the naked eye which observed that the fixtured slide clamp component was assembled backward.A functional testing was performed including clear passage and pressure testing; and the device performed according to product specifications.The reported condition was verified during initial inspection.The cause of the condition was due incorrect assembly during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SOLUTION ADMINISTRATION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11486495
Report Number1416980-2021-01331
Device Sequence Number3412328
Product Code FPA
UDI-Device Identifier00085412015477
UDI-Public(01)00085412015477
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K870940
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number1W5000
Device Lot NumberDR20H13069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2021
Initial Date Received by Manufacturer 02/16/2021
Supplement Date Received by Manufacturer04/15/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date04/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALBUMIN; UNSPECIFIED INFUSION PUMP
-
-