TELEFLEX MEDICAL RUSCH MRI SET: HDL+ MIL 1,2,3 MAC 2,3,4; LARYNGOSCOPE, RIGID Back to Search Results
Model Number IPN048624
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
It was reported "client used mri kit 005852300 and blade jammed on the handle and cannot be removed".No patient involvement reported.
 
Event or Problem Description
It was reported "client used mri kit 005852300 and blade jammed on the handle and cannot be removed".No patient involvement reported.
 
Additional Manufacturer Narrative
(b)(4), the customer returned one rusch mri set (005852300) containing 6 varying blades (mil 1, mil 2, mil 3 , mac 2, mac 3, mac 4) and one handle.The components were visually inspected for signs of abuse/misuse/damage and nothing was found.The mac 2 blade was returned on the handle.The blades in the kit were functionally tested per ifu l07085 rev.01 which states, "attach blade to handle.Click in place.To switch on, pull blade up.To switch off, move blade down." the mac 2 blade was not able to be removed from the returned handle.The other five other blades were able to be connected to a lab inventory rusch mri handle, engaged, and disengaged with no issues.R and d was consulted about the mac 2 blade stuck on the mri handle.They were able to remove the blade from the handle and found that both components passed the relevant dimensional inspections.They also found that the handle had sharp edges and deformation where the blade's ball bearings would engage.It is unknown what caused this damage.The device history record of lot 200801 was reviewed and no issue that could have contributed to the reported failure was noted.The device was manufactured according to release specification.The instructions for use (ifu) provided with this kit instructs the user, "attach blade to handle.Click in place.To switch on, pull blade up.To switch off, move blade down." the complaint was confirmed via functional inspection.The returned mac 2 blade was found to be stuck on the mri handle.Once the handle was removed from the blade, the handle was found to have sharp edges and deformation where the blade's ball bearings would engage.The root cause of this failure is unknown.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUSCH MRI SET: HDL+ MIL 1,2,3 MAC 2,3,4
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
MDR Report Key11487299
Report Number8030121-2021-00010
Device Sequence Number3678641
Product Code CCW
UDI-Device Identifier14026704616371
UDI-Public14026704616371
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberIPN048624
Device Catalogue Number005852300
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2021
Initial Date Received by Manufacturer 02/24/2021
Supplement Date Received by Manufacturer04/15/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date04/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-