| Model Number |
2.1 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Change healthcare investigated the reported issue and determined there was a software defect in change healthcare enterprise viewer (chev).The problem occurs when a scheduled study exists in a pacs patient record and a more recent performed study from the same patient record is opened in chev.Images from the opened (anchor) study are displayed in the viewport but the anchor study details and image thumbnails are not displayed on the patient timeline.The missing timeline information could potentially lead to a study context error.
|
| |
|
Event or Problem Description
|
|
A neurosurgeon initially misinterpreted anchor images displayed in the viewport of the change healthcare enterprise viewer (chev) as a prior study.It was reported this caused the neurosurgeon to almost use the wrong image for surgery.No adverse consequences were reported.
|
| |
|