SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122546
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Necrosis (1971); Pain (1994); Metal Related Pathology (4530)
Date of Event 04/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that a revision surgery was performed on the patients right hip on (b)(6) 2019 due to a painful metal-on-metal right total hip replacement with a large pseudotumor, there was a significant amount of periarticular necrosis of the soft tissue with approximately 50-75% compromise of the abductor muscles.The cup and the stem were retained.
 
Additional Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the r3 shell, r3 liner, hemi head, sleeve and stem was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the r3 shell, r3 liner, hemi head and stem.Similar complaints have been identified for the sleeve.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product's instructions for use found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The reported elevated cocr levels, and presence of a large pseudotumor, and soft tissue and muscle necrosis were observed intraoperatively during the revision of rt mom tha, which was in situ for 9 years.These intraoperative findings are consistent with findings associated with metal debris.However, the clinical root cause cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Additional Manufacturer Narrative
H10: additional information in a2, a3, b6 and b7.
 
Additional Manufacturer Narrative
H3, h6: it was reported that right hip revision surgery was performed due to pain.Pseudotumor and necrosis were observed intraoperatively.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the historical complaints data for the devices concerned was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the hemi head and the modular sleeve.However, as the devices are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The newly added medical records were reviewed.The reported elevated cocr levels, and presence of a large pseudotumor, and soft tissue and muscle necrosis were observed intraoperatively during the revision of rt mom tha, which was in situ for 9 years.These intraoperative findings are consistent with findings associated with metal debris.The report of the patient having a ¿limited lumbar spine which affects his spinal pelvic motion and compounds the instability issues¿ and the cup orientation being ¿in more anteversion than ideal" cannot be ruled out as contributing factors of the subsequent multiple dislocations, which ultimately led to the 2nd revision.However, without x-ray images this cannot be confirmed.Therefore, the clinical root cause cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 46MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11489837
Report Number3005975929-2021-00147
Device Sequence Number19832788
Product Code NXT
UDI-Device Identifier00885556071588
UDI-Public885556071588
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2019
Reporter Type Manufacturer
Report Source Other,Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/30/2013
Device Catalogue Number74122546
Device Lot Number08GW18076
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 02/18/2021
Supplement Date Received by Manufacturer04/28/2021
05/25/2022
09/15/2022
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date05/10/2021
06/01/2022
09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/31/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SLEEVE 74222300, LOT 09LW25119.; SLEEVE 74222300, LOT 09LW25119
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexMale
-
-