Model Number 7211005
Medical Device Problem Codes Contamination /Decontamination Problem (2895); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that when the dyonics 25 inflow/outflow tube set package was opened, the surgeon noticed that the cellophane tape was already attached to inside the sterilization tray and a foreign material similar to dust was found to be sealed in a package, so the staff taped the foreign material to the area where it was found.The procedure was finished with the same device and there was no delay nor other complication reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device noted there was a piece of the shipping container made of corrugated fiberboard inside the sterile packaging.Functional evaluation was not performed as visual inspection has confirmed the complaint.The complaint has been confirmed.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include transportation and handling of the packaging.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of the manufacturing process shows that there are precautions taken to eliminate particulates inside the sterile packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYONICS 25 INFLOW/OUTFLOW TUBE SET (3)
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11490015
Report Number1643264-2021-01023
Device Sequence Number5424864
Product Code HRX
UDI-Device Identifier03596010543578
UDI-Public03596010543578
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2023
Device Model Number7211005
Device Catalogue Number7211005
Device Lot Number50885218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2021
Initial Date Received by Manufacturer 02/19/2021
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date03/15/2021
Supplement Report FDA Received Date04/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-