| Model Number |
TMTB11 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Purulent Discharge (1812); Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
|
| Date of Event |
02/06/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet (b)(4).Patient weight is not provided / unknown.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
Doctor reports that the tmtb11 implant was removed due to peri-implantitis.Heavy amounts of purulence emerged from the implant socket.Socket was debrided and irrigated and grafted.Tooth site # 7.Pain was a consequence of the peri-implantitis.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: type of investigation was added: 3331.A device history record (dhr) review was completed for the reported device lot.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance, which could have caused or contributed to the reported event was noted as part of the dhr.A sterilization record review was completed and all sterilization activities were conducted with no nonconformities.A complaint history review was completed for the reported device lot for similar event and no other complaint was identified.A definitive root cause for the reported event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Search Alerts/Recalls
|