• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 5000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 5000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938743
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a 5000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked from the additive port during filling.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The lot was manufactured from august 07, 2018 - august 08, 2018.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak between the spike port tube and the spike port bonding area.The reported condition was verified.The cause of the condition could not be determined; however, the most likely cause was due to inadequate or lack of cyclohexanone applied to the spike port cap tube when it was inserted to the spike port during the manufacturing process causing an incorrect bonding.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5000 ML TPN BAG
Common Device Name
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11494452
Report Number1416980-2021-01353
Device Sequence Number17139983
Product Code LHI
UDI-Device Identifier00085412477275
UDI-Public(01)00085412477275
Combination Product (Y/N)Y
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K900585
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2021
Device Catalogue NumberH938743
Device Lot Number60139369
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2021
Initial Date Received by Manufacturer 02/17/2021
Supplement Date Received by Manufacturer03/30/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date04/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-