| Model Number |
60NFP35 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer facility phone number: (b)(6).Device evaluation in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a tiny hole in the bottom of the flextend portion of the bivona tracheostomy tube.This was near the outer hub, and was exposing the viewing of the wire in the flextend portion.This product fault was noticed after a whistling sound was heard coming from the child.The ventilator, which was also on, was alarming.The patient's mother tried to suction the child's tube but she was unable to insert the suction catheter to the needed suction depth.The mother thought that there was a blockage.After suspecting this, the mother performed a tracheostomy tube change.No patient injury, or further complications were reported in relation to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: , corrected data: patient age corrected in section a.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: one sample was returned and a tube was found to be crushed in the same place a cut in the silicon was found.Based on the conditions of sample received i.E.Used and crushed on surface tube.The most probable root cause is that the unit became damaged after left smiths medical facilities.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
| |
|