| Model Number |
70914-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the freestyle freedom lite meter were reviewed and the dhrs showed the freestyle freedom lite meter passed all tests prior to release.Dhrs for the freestyle test strips were reviewed and the dhrs showed the freestyle test strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.If the product is returned, the case will be re-opened and a physical investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 65 mg/dl, 200 mg/dl and 40 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Meter (b)(6) has been returned, and investigated with returned test strips.Visual inspection has been performed on the returned meter, and no issues were observed.Meter powered on with button and with strip insertion.Control solution testing was performed, and no issues were observed.All results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported, receiving erratic readings on their adc blood glucose meter.Customer reported, receiving readings of 65 mg/dl, 200 mg/dl and 40 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Search Alerts/Recalls
|