• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000254
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the intra-aortic balloon pump (iabp) was alarming for a helium leak, and blood noted in the intra-aortic balloon (iab).The balloon was clamped and disconnected.As a result, the doctor was notified and the iab was removed.There was no report of patient complications, serious injury or death.
 
Manufacturer Narrative
(b)(4).No part was returned to teleflex chelmsford for investigation.The reported complaint of iab blood in helium pathway is not able to be confirmed.The root cause of the complaint is undetermined.The specific serial number/lot number was not reported, but a device history record (dhr) review was conducted for the lot numbers shipped to this account with no relevant findings.All devices passed manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
 
Event Description
It was reported that the intra-aortic balloon pump (iabp) was alarming for a helium leak, and blood noted in the intra-aortic balloon (iab).The balloon was clamped and disconnected.As a result, the doctor was notified and the iab was removed.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBEROPTIX ULTRA 8 IAB: 8FR 40CC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key11494940
MDR Text Key240369616
Report Number3010532612-2021-00075
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902007247
UDI-Public00801902007247
Combination Product (y/n)N
PMA/PMN Number
K021462
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN000254
Device Catalogue NumberIAB-05840-LWS
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/24/2021
Initial Date FDA Received03/16/2021
Supplement Dates Manufacturer Received04/20/2021
Supplement Dates FDA Received04/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-