ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC
|
Back to Search Results |
|
Model Number IPN000254 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the intra-aortic balloon pump (iabp) was alarming for a helium leak, and blood noted in the intra-aortic balloon (iab).The balloon was clamped and disconnected.As a result, the doctor was notified and the iab was removed.There was no report of patient complications, serious injury or death.
|
|
Manufacturer Narrative
|
(b)(4).No part was returned to teleflex chelmsford for investigation.The reported complaint of iab blood in helium pathway is not able to be confirmed.The root cause of the complaint is undetermined.The specific serial number/lot number was not reported, but a device history record (dhr) review was conducted for the lot numbers shipped to this account with no relevant findings.All devices passed manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
|
Event Description
|
It was reported that the intra-aortic balloon pump (iabp) was alarming for a helium leak, and blood noted in the intra-aortic balloon (iab).The balloon was clamped and disconnected.As a result, the doctor was notified and the iab was removed.There was no report of patient complications, serious injury or death.
|
|